Focus – AirSep MN172-1 User Manual
Page 218

EN
Focus™
EN-30
AirSep
®
Corporation
MN172-1 rev. A - 02/13
Status
Audible Alarm
Light
Indicates
Action
Alarm
Intermittent:
Beep, beep, beep
(Amber) alarm;
intermittent light
Breathing rate is
exceeding the
capacity of the
Focus unit.
Reduce activity, and
then if necessary use
another source of
oxygen, as available.
Contact your
Equipment Provider.
Alarm
Audible alarm
sounds
(beep,beep)
Audible alarm
sounds
(beep,beep,beep)
(Red) light
General
malfunction of the
Focus unit has
occurred.
Turn off the unit.
Change to another
source of oxygen,
and contact your
Equipment Provider.
Check air inlet and
exhaust vents for
obstruction.
Clear obstruction at
air inlet and exhaust.
See also other documents in the category AirSep Health and hygiene:
- MN170-1 D (40 pages)
- MN189-1 (12 pages)
- MN123-1 F (45 pages)
- MN131-1 B (556 pages)
- MN166-1 B (40 pages)
- MN167-1 (520 pages)
- MN168-1 (3 pages)
- MN169-1 (40 pages)
- MN092-1 D (13 pages)
- MN104-1 D (40 pages)
- MN112-1 (208 pages)
- MN119-1 D (4 pages)
- MN120-1 (44 pages)
- MN078-1 F (42 pages)
- MN084-1 D (137 pages)
- MN139-1 (144 pages)
- MN164-1 (30 pages)
- MN152-1 (18 pages)
- MN136-1 H (27 pages)
- MN137-1 D (190 pages)
- MN156-1 B (18 pages)
- MN157-1 B (332 pages)
- MN165-1 G (28 pages)
- MN175-1 (2 pages)
- 15062781 (16 pages)
- 4111-7 (95 pages)
- 20626704 (24 pages)
- 20626704-C1 (154 pages)
- ML-CONC0056 (2 pages)
- MN028-1 (1 page)
- MN171-1 (23 pages)
- MN083-1 (1 page)
- MN126-1 C (69 pages)
- MN179-1 B (62 pages)
- MN110-1 (30 pages)
- MN105-1 C (52 pages)
- MN118-1 B (40 pages)
- MN134-1 B (40 pages)
- MN151-1 (1 page)
- MN188-1 (1 page)
- MN188-2 (2 pages)
- MN155-1 (16 pages)
- MN138-1 H (46 pages)
- MN178-1 (1 page)