beautypg.com

Graph tab, 7 graph tab – PNI TCM User Manual

Page 39

background image

PNI Sensor Corporation

DOC#1014688 r06

TCM User Manual

– July 2013

Page 34

6.7

Graph Tab

The graph provides a 2-axis (X,Y) plot of the measured field strength. If “w/o User Cal”

graph data is selected, the plot and data provide magnetic field strength measurements after
the FIR taps are applied, but prior to applying the user calibration coefficients. If “with User
Cal” graph data is selected, the plot and data provide data after applying the FIR filter and the
user calibration coefficients. The sample plot shows a 360° rotation in the horizontal plane,
with both “w/o User Cal” and “with User Cal” selected. The offset between these two plots
represents the effect of the calibration coefficients. The graph can be used to visually see

hard and soft iron effects within the environment measured by the TCM, as well as corrected

output after a user calibration has been performed.