Note: While the TCM is shown removed from the host system, the Accelerometer
Calibration may be performed with the TCM mounted in the host system.

Figure 5-2: Accelerometer Calibration Starting Orientations

5.2.2

Mag-and-Accel Calibration

The TCM allows for a simultaneous magnetometer and accelerometer calibration. This

requires a full-coverage calibration pattern, physically stable measurements, and
installation in the user’s system so the host system’s magnetic signature is present. PNI
recommends 18 to 32 calibration points for a Mag-and-Accel Calibration. The

Accelerometer-Only Calibration pattern discussed in Section 5.2 will work for a Mag-

and-Accel Calibration. Optimal performance is obtained when all rotations of the TCM

are performed towards magnetic north to achieve the widest possible magnetic field

distribution.

Note that combining calibrations only makes sense if all the host system’s magnetic

distortions (steel structures or batteries, for instance) are present and fixed relative to the

module when calibrating. If an Accelerometer-Only Calibration is performed, the user’s

system distortions are not relevant, which allows the TCM to be removed from the host

system in order to perform the Accelerometer-Only Calibration.