beautypg.com

Hpr during calibration, Calibration settings – PNI TCM User Manual

Page 31

background image

PNI Sensor Corporation

DOC#1014688 r06

TCM User Manual

– July 2013

Page 26

Acquire Delay

The Acquire Delay sets the time between samples taken by the module, in seconds.

This is an internal setting that is NOT tied to the time with which the module

transmits data to TCM Studio or the host system. Generally speaking, the Acquire

Delay is either set to 0, in which case the TCM is constantly sampling or set to equal

either the Poll Delay or Sample Delay values. The advantage of running with an

Acquire Delay of 0 is that the FIR filter can run with a relatively high Tap value to

provide stable and timely data. The advantage of using a greater Acquire Delay is

that power consumption can be reduced, assuming the Sample or Poll Delay are no

less than the Acquire Delay.

Sample Delay

The Sample Delay is relevant when Continuous Mode is selected. It is the time

delay, in seconds, between completion of the TCM sending one set of data and the

start of sending the next sample set. If the delay is set to 0, then the TCM will begin

sending new data as soon as the previous data set has been sent. Note that the inverse

of the Sample Delay is greater than the sample rate, since the Sample Delay does not

include the actual measurement acquisition time. The default is 0.

Flush Filters

Flushing the FIR filter clears all the filter values so it is necessary to fully repopulate

the filter before a good reading can be given. For example, if 32 FIR taps is set, then

32 new samples must be taken to provide a good reading. It is particularly prudent to

flush the filter if the Sample Delay is set to a non-zero value as this will purge old

data. Note that flushing the filters increases the delay until data is output, with the

length of the delay being directly correlated to the number of FIR taps. The default is

not to Flush Filters.

6.3.8

HPR During Calibration

When the button is selected, heading, pitch, and roll will be output on the

Calibration tab during a calibration.

6.3.9

Calibration Settings

Automatic Sampling

When selected, the module will take a sample point once the minimum change and

stability requirements have been satisfied. If the user wants to have more control

over when the point will be taken, then Auto Sampling should be deselected. Once

deselected, the button on the Calibration tab will be active. Selecting