Limited tilt range calibration, Table 5-3: 12 point 2d calibration pattern, 3 limited tilt range calibration – PNI TCM User Manual

Page 22

background image

PNI Sensor Corporation

DOC#1014688 r06

TCM User Manual

– July 2013

Page 17

Table 5-3: 12 Point 2D Calibration Pattern

Sample #

Yaw

Pitch

1

Roll

1

1

2

30°

max. negative

max. negative

3

60°

4

90°

max. positive

max. positive

5

120°

6

150°

max. negative

max. negative

7

180°

8

210°

max. positive

max. positive

9

240°

10

270°

max. negative

max. negative

11

300°

12

330°

max. positive

max. positive

Footnote:

1. For best results, the tilt experienced during calibration should match that experienced

in service. For example, if the TCM is restrained to a level plane in service, then
calibration should be

in a plane, where “max. positive” and “max. negative” are 0°.

5.1.3

Limited Tilt Range Calibration

A Limited Tilt Range Calibration is recommended when 45

° of tilt isn’t feasible, but >5°

of tilt is possible. It provides both hard-iron and softiron distortion correction. The

recommended 12 point calibration pattern given below is a series of 3 circles of evenly

spaced points, with as much tilt variation as expected during use.

Table 5-4: 12 Point Limited-Tilt Calibration Pattern

Sample #

Yaw

Pitch

Roll

First Circle

1

2

90°

3

180°

6

270°

Second Circle

7

45°

> +5°

> +5°

8

135°

> +5°

> +5°

11

225°

> +5°

> +5°

12

315°

> +5°

> +5°

Third Circle

13

45°

< -5°

< -5°

14

135°

< -5°

< -5°

17

225°

< -5°

< -5°

18

315°

< -5°

< -5°