beautypg.com

Accelerometer-only calibration – PNI TCM User Manual

Page 24

background image

PNI Sensor Corporation

DOC#1014688 r06

TCM User Manual

– July 2013

Page 19

vector, which in turn will result in an inaccurate heading reading. For this reason, the TCM

should be stationary when taking a measurement.

As previously mentioned, PNI calibrates the accelerometer in its factory prior to shipment.

But over time the bias and offset of the accelerometer will drift. For this reason PNI

recommends the accelerometer be recalibrated every 6 to 12 months. The user may return

the TCM to PNI for accelerometer calibration, or the user may perform a user accelerometer

calibration. The remainder of this section covers the user accelerometer calibration.

5.2.1

Accelerometer-Only Calibration

The requirements for a good user accelerometer calibration differ significantly from the

requirements for a good magnetic calibration. Specifically, a good accelerometer

calibration involves the TCM experiencing a wide range of pitch and roll values,

preferably seeing both ±180° of pitch and ±180° of roll. Also, it is necessary for the

TCM to be very still during an accelerometer calibration. If possible, PNI recommends

using a fixture to hold the device during calibration, although resting the TCM on a hard

surface normally is sufficient.

The accelerometer either can be calibrated while mounted in the host system or it may be

removed and calibrated outside the system. The advantage of performing the calibration

while mounted in the host system is the user does not need to remove the TCM from the

system, which can be burdensome, and a simultaneous Mag-and-Accel Calibration may

be appropriate. The advantage of performing the calibration outside of the system is it

may be much simpler to obtain the desired range of pitch and roll.

Figure 5-2 shows the two basic starting positions for the recommended 18-point

calibration pattern. Starting with the TCM as shown on the left in Figure 5-2, rotate the

device about its z axis such that it sits on each of its 4 edges, taking one calibration

sample on each edge. Then place the TCM flat on the surface and take a calibration

sample, then flip it over (roll it 180°) and take another sample. Next, starting with the

TCM as shown on the right, take a calibration point with it being vertical (0°). Now tilt

the TCM back 45° and take another calibration point (+45°), then tilt the device forward

45° and take another calibration point (-45°). Repeat this 3-point calibration process for

the TCM with it resting on each of its 4 corners. Note that it is possible to perform an

Accelerometer Calibration with as few as 12 sample points, although it generally is more

difficult to obtain a good calibration with just 12 sample points. Also, the maximum

number of calibration points is 18.