FUTEK IPM52x (D520) Digital Display User Manual
Page 29

28
"CHAn 1", "CHAn 2". "CHAn 3" Select channel to do
peak on
4. "Scale 2" setup
Programs the setup and assignment of the second scale
factor. This would typically be used to switch the display
between English and Metric engineering units. The
selection of scale 2 can be activated from a Front Panel
Key (F1 - F4) or a remote input. They must be programmed
to perform the switch to scale 2 under the "SPEC"
branch programming.
"CHAn 1", "CHAn 2" Select which channel the second
scale factor is being setup for. Both channel 1 and channel
2 can be setup for an individual second scale factor.
"dEC Pt" selects decimal point. Push SELECT key for
displaying decimal point selections and ENTER key to
confirm choice.
"In Lo" enter in low input value
"dSP Lo" (display low) enter in display value
corresponding to low input value.
"In HI" enter in high input value.
"dSP Hi" (display high) enter in display value
corresponding to high input value.
5. LOCK OUT CODE SETUP
This enable the program of a lockout code (0 to 99) to limit
the entry into the setup mode. Once a code is entered here,
it must be enter upon trying to enter the
setup mode when the meter prompts for "CODE". Setting
the code to 00 disable the lockout code function.
"CodE" is flashed with the lock out code currently
programmed. Enter in a new code or exit the program
(EXIT key).
6. DEFAULT program
This program is used for any of the following:
• Reset the meter back to the factory defaults.
• Set the user defined defaults into memory