FUTEK IPM52x (D520) Digital Display User Manual
Page 20
19
high input values below. Push ENTER key to
advance.
2. "dEC Pt" selects decimal point. Push SELECT
key for displaying decimal point selections and
ENTER key to confirm choice.
3. "In Lo" enter in low input value
4. "dSP Lo" (display low) enter in display value
corresponding to low input value.
5. "In HI" enter in high input value.
6. "dSP Hi" (display high) enter in display value
corresponding to high input value.
b. Method 2 (Ld CAL)
NOTE: Input Signal must be applied to the unit for low
and high display values. The meter compares the input
signal to the display value entered and scales the meter
based on the two inputs.
1. Raw input is displayed. This can be used to
make sure the correct input is being applied
and/or that the meter is properly measuring it.
Push enter to advance.
2. "dEC Pt" selects decimal point. Push SELECT
key for displaying decimal point selections and
ENTER key to confirm choice.
3. "dSP Lo" (display low).
Either select "Yes" or "nO" here.
"Yes" allows you to apply corresponding low
input signal to input and enter in corresponding
low display value. Push ENTER key to make
displayed value equal input. If “NO” is selected
it skips over this program and leaves values to
what they were.