Figure 20: hardware reset, Figure 20: hardware reset 26, Normally open pushbutton switch – Badger Meter Model IOG User Manual
Page 26

Clear Grand Total (Clr G-T)
Basic Function
At the clear grand total Clr G-T prompt, press the ENTER button once The display now says no on the screen To clear the
grand total press either the
▲ arrow or the ► arrow button to change from No to Yes Press the ENTER button to select Yes
and advance to the next parameter The totalizer can also be reset using a hardware reset as shown in
JP1
JP2
JP3
Input
Total P
ulse
Sig
nal
P1
Freq. In
4-20mA
Iso Total Pluse
TR_B
TR_A
RS485 Gnd
Setpoint 1
Setpoint 2
Gnd
+
–
+
–
+
–
Total Reset
OC Total Pluse
Signal Gnd
TB1
Mag
Pulse
Iso
OC
Low
High
Normally Open
Pushbutton
Switch
Figure 20: Hardware reset
Password (Passwd)
Basic Function
Password protection prevents unauthorized users from changing programming information Initially, the password is set to all
zeros If the current setting requires a change, press the ENTER button once at the password Passwd prompt The first digit of
the password begins to flash Use the
▲ arrow button to increment the digit until it matches the first number of the desired
password Use the
► arrow button to advance to the next digit of the desired password value then use the ▲ arrow button
to increment the display digit until it matches the next digit of the desired password Repeat this step for the all the digits of
the password and then press the ENTER button once to save the new password and advance to the next parameter
Reset Password (RstPswd)
Basic Function
Reset password parameter prevents unauthorized users from manually resetting the flow monitor’s main totalizer Initially,
the password is set to all zeros To change the reset password, press the ENTER button once at the reset password RstPswd
prompt The first digit of the password begins to flash Use the
▲ arrow button to increment the digit until it matches the first
number of the desired password Use the
► arrow button to advance to the next digit of the desired password value then
use the
▲ arrow button to increment the display digit until it matches the next digit of the desired password Repeat this
step for the all the digits of the password and then press the ENTER button once to save the new password and advance to
the next parameter
NOTE:
N
Entering a password in the Passwd screen and leaving the password blank in the RstPswd screen allows for total resets
(not requiring a password), but restricts programming modification
ER-500 Flow Monitor
Page 26
August 2013