Thermo Fisher Scientific EcoScan DO 6 User Manual

Page 69

background image

Instruction Manual

DO 6

61

% Saturation Calibration Limits

Factory calibrated absolute value of 10.0%
and below for 0% point & 50% to 200% for
100% point.

Concentration Calibration Window

+/- 40% from the factory default
measurement value. Minimum reading
allowed is 2.00 mg/L (ppm).

Temperature Calibration Window

+/- 5°C from factory default measurement

Offset Adjustments (% Saturation)

+/- 10.0 of reading in Saturation mode

HOLD function

Yes

Auto-Off function

Selectable Auto Off function. (20 minutes
after last key press)

Display

Customs Single 4 Digit LCD

Inputs

BNC for DO & 2.5 mm Phono for
temperature

Operating Range

0 to 50 °C

Power Requirements

4 AAA-sized batteries (included)

Battery Life

> 700 hours (Alkaline Batteries)

Dimensions

Meter: 14 cm (L) x 7 cm (W) x 3.5 cm (H)

Probe: 115 mm (L) x 12 mm (Dia) with 3-ft
cable.

Probe’s membrane housing: 16 mm (Dia)