7.2 (COF.1)

Configuration

Menu

In this menu, the meter lets you configure the
selection of automatic temperature
compensation (A) and the selection of mg/L
or ppm mode for the mg/L (ppm)
Concentration measurement mode. (Second
feature is only available in mg/L (ppm)
Concentration Set Up menu)

7.2.1

Selection of Automatic
Temperature Compensation

From “COF.1” of any Set Up menu,

1) Press the HOLD/ENTER key for the

display to show the Automatic
Temperature Compensation selection
menu “A.ATC”.

2) Press the HOLD/ENTER key to enter

the ATC selection menu.

3) Use

the

▲ and ▼ keys to either activate

or deactivate the ATC feature.

a. YES – activates the ATC

b. NO – deactivates the ATC and

activates the Manual temperature
compensation

4) Press the HOLD/ENTER key to confirm

your selection.

HOLD

ENTER

HOLD

ENTER

Figure 20: Selection of Automatic or

Manual Temperature

Compensation