DO 6

29

5.3.2 Salinity

Setting

Adjustment

NOTE: This mode is available only from
the mg/L (ppm) Concentration Set Up
menu.

1) Repeat steps 1 to 5 as in the steps of

Pressure Setting Adjustment.

2) Press the HOLD/ENTER key twice till

the meter displays the Salinity setting
mode “b.SAL”.

3) Press the HOLD/ENTER key to enter

the Salinity setting mode.

4) Use

the

▲ and ▼ keys to enter the

salinity of your solution in parts per
thousand (ppt). Factory default is 0.0.
Salinity values up to 50 ppt can be
entered.

5) Press HOLD/ENTER key to confirm

the value. A confirmation indicator
“CO” will be displayed for 1.5 seconds
before returning to the main group
menu “DPr.7”.

6) Press CAL key to return to the

measurement mode.

Figure 16: Salinity Setting Adjustment Sequence

Display will return back to “DPr.7”
subgroup menu