7 (ofs.6) % saturation offset adjustment – Thermo Fisher Scientific EcoScan DO 6 User Manual

Page 55

background image

Instruction Manual

DO 6

47

7.7

(OFS.6) % Saturation
Offset Adjustment

This is a useful feature that allows
you to offset meter’s value when
cross referenced with another DO
meter. That way, it can be
standardized without you having to
perform manual calculation. Your DO
6 meter allows % Saturation Offset
adjustment within +/- 10.0% offset
and its adjusted offset value can be
viewed in the Electrode Properties
menu “ELE.3”.

From the % Saturation measurement
mode,

1) Dip the DO electrode in the

sample solution and allow the
reading to stabilize.

2) Check the reading of another

DO meter being used as a
reference. This reference meter
should have its probe immersed
in the same sample solution and
at the same depth.

3) Switch off your meter and enter

the % Saturation Set Up menu.

4) Scroll the Set Up main group

menu by using the ▲ and ▼
keys until the meter displays the
% Saturation Offset Adjustment
menu “OFS.6”.

5) Press the HOLD/ENTER key to

enter the menu. The display will

Figure 27: % Saturation Offset

Adjustment

%

HOLD

ENTER

Measured reading based on last
percentage saturation calibration
is displayed momentarily

%

%

HOLD

ENTER

Last offset adjusted

value will be displayed

Display will return to “OFS.6” subgroup menu

“CO” will display for
1.5 seconds