DO 6

20

Figure 11: mg/L (ppm)

Concentration Calibration Sequence

4.5 Dissolved

Oxygen

Calibration in mg/L or
ppm concentration mode

Calibrating the meter to 100% Saturation
will also calibrate the concentration mode
at the value in mg/L corresponding to
100% Saturation. This should produce
acceptable results in most applications.

This meter also lets you make a calibration
adjustment in mg/L (ppm) Concentration
mode without affecting your % Saturation
calibration.

To select between mg/L and ppm units,
see page 38, “Selection of mg/L or ppm
mode”.

4.5.1

To calibrate in mg/L (ppm)
Concentration mode:

1) Calibrate 100% Saturation as per

Section 4.4.

2) Rinse the probe well with DI water or

rinse solution. Wipe the outside of the
probe carefully (do not touch the
membrane).

3) Dip the probe into a sample of known

oxygen concentration. Wait for the
reading to stabilize.

4) Press the MODE key to select the

mg/L (ppm) Concentration mode.

5) Press the CAL key. The display will

show a value which is with respect to

mg/l

mg/l

mg/l

CAL

CAL

HOLD

ENTER

CAL

mg/l

“CO” will display

for 1.5 seconds

A

A

A value with respect to the
factory default calibration
will be displayed