Thermo Fisher Scientific EcoScan DO 6 User Manual

Page 59

background image

Instruction Manual

DO 6

51

7.8.2 Salinity

Setting

Adjustment

Note: This mode is available only from the
mg/L (ppm) Concentration Set Up menu.

1) Repeat steps 1 to 5 as in the steps of

Pressure Setting Adjustment.

2) Press the HOLD/ENTER key twice till

the meter displays the Salinity setting
mode “b.SAL”.

3) Press the HOLD/ENTER key to enter

the Salinity setting mode.

4) Use

the

▲ and ▼ keys to enter the

salinity of your solution in ppt.

Note: The salinity value is in ppt (parts per
thousand)

5) Press HOLD/ENTER key to confirm

the value. A confirmation indicator
“CO” will be displayed for 1.5 seconds
before returning to the subgroup
menu “DPr.7”.

6) Press CAL key to return to the

measurement mode.

Display will return back to “DPr.7”
subgroup menu

HOLD

ENTER

HOLD

ENTER

“CO” will display for

1.5 seconds

Figure 29: Salinity Setting

Adjustment Sequence