Instruction Manual

DO 6

44

7.6

(rSt.5) Reset to Factory Default

This mode lets you reset all parameters to
factory default settings. There are two
levels of reset:

1) Calibration Reset: - Reset only the

calibration values. This clears all
previous calibrated values and reset
to factory default. See Section 11 for
factory default settings.

2) User Reset: - Clears all data including

calibration and other customized
setup functions and reset it to factory
default. See Section 11 for factory
default settings.

7.6.1

Calibration Reset

From any Set Up menu:

1) Scroll the Set Up main group menu

by using the ▲ and ▼ keys until the
meter displays the Reset to Factory
Default menu “rSt.5”.

2) Press the HOLD/ENTER key to enter

the menu.

3) Press the HOLD/ENTER key to enter

the Calibration Reset menu “A.CAL”.

4) Use

the

▲ and ▼ keys to activate or

deactivate the Calibration Reset
selection.

a. YES – activates Calibration

Reset

Figure 25: Calibration reset

selection

HOLD

ENTER

HOLD

ENTER

Calibration
Reset