10 ERROR

MESSAGES

LCD Display

Indicates Cause

Solution

Low
Battery
indicator

Low battery
level.

Need new batteries or
battery connection is
bad.

Clean battery contacts.
Replace batteries with
fresh ones, noting
polarity.

“Err 1” in %
Saturation Mode

% Saturation
Calibration
error

Calibration is
performed when factory
calibrated absolute
value is out of
calibration range –
10.1% to 49.9%

Check the value of the
calibration solution. If
zero calibration is done,
make sure the limit of
10% is not exceeded.
If message persist,
recondition your probe.

“Err. 1” in mg/L
(ppm)
Concentration
Mode

Concentration
Calibration
error

Calibration is
performed when the
calibration solution
factory absolute
measurement is below
2.00 mg/L or ppm

Check the calibrating
solution to be above 2
mg/L or ppm.

Check that the correct
temperature and salinity
setting has been set
prior calibration.

“UR”/”OR” with
blinking “A”
annunciator in
Temperature
Mode

ATC probe
error, Under
Range, Over
Range

ATC probe is removed
or broken while the
meter has the ATC
feature activated.

Check the probe’s
temperature input
phono jack connection
to the meter. Ensure
probe is not broken or
punctured.

“----“with blinking
“A” annunciator
in % Saturation
and
Concentration
Mode

ATC probe
error

ATC probe is removed
or broken while the
meter has the ATC
feature activated.

Check the probe’s
temperature input
phono jack connection
to the meter. Ensure
probe is not broken or
punctured.