Thermo Fisher Scientific EcoScan DO 6 User Manual

Page 31

background image

Instruction Manual

DO 6

23

5.2 Manual

Temperature

Compensation

For

manual

temperature

compensation, simply deactivate the
“A.ATC” mode as explained in the set
up menu on page 36. The automatic
temperature compensation annunciator
“A” will not be displayed to represent
that the meter is in manual temperature
compensation mode.

To use manual temperature
compensation, you need to enter the
temperature value of your process into
the meter. You can select any
temperature between 0 and 50 °C.
Default value is 25 °C.

1) Switch on the meter and if

necessary, press MODE key to
select temperature measurement
mode.

2) Press CAL key to start temperature

calibration process.

3) The “CAL” indicator will start

blinking and the display will show
the last manual set temperature
value.

4) Check the temperature of your

sample using an accurate
reference thermometer. Wait for
the thermometer value to stabilize.
Press the ▲ or ▼ key of your
meter to manually set its
temperature value according to the
value of the reference thermometer
used.

CAL

CAL

T

T

T

T

CAL

HOLD

ENTER

Last set manual
temperature setting

“CO” will display for 1.5

seconds

Figure 13: Manual Temperature

Compensation value selection sequence