Instruction Manual

DO 6

41

7.4.3

Viewing the 100% Saturation mV value

This parameter shows the sensor’s mV output corresponding to 100%
Saturation.

From Step 5 above,

6) Press the HOLD/ENTER key. The display will show “HI.mV” momentarily

before displaying the sensor’s mV output value with respect to 100%
Saturation.

7.4.4

Viewing the 0% Saturation mV value

This parameter lets you view the sensor’s mV output corresponding to 0%
Saturation.

From Step 6 above,

7) Press the HOLD/ENTER key. The display will show “LO.mV” momentarily

before displaying the sensor’s mV output value with respect to 0%
Saturation.

8) Press the HOLD/ENTER key again to exit to the main group menu. Press

CAL to return to the measurement mode.

Note: To view the electrode properties in the mg/L (ppm) Concentration mode,
repeat the whole steps above using the mg/L (ppm) Concentration Set Up
menu. Viewing of the % Saturation Offset will be skipped in this menu.