DO 6

16

4) Press

the

▲ and ▼ keys to adjust the meter reading to the correct

temperature value (i.e. the temperature of the temperature bath)

5) Press the HOLD/ENTER key to confirm temperature calibration and return

to measurement mode.

NOTE:

To exit from Temperature Calibration mode without confirming the calibration,
press CAL instead of HOLD/ENTER.

Temperature calibration is restricted to ± 5°C from the factory default
temperature measurement displayed during calibration.