ADVANCED SETUP FUNCTIONS

7.1

Advanced Setup Overview

The advanced setup mode lets you customize your meter’s preferences and
defaults. There are two Set Up menus: - one which is derived from the %
Saturation or Temperature mode and the other is derived from the mg/L (ppm)
Concentration mode.

7.1.1

To enter the % Saturation or Temperature Set Up menu:

1) Make sure that the meter is switched-off while in the % Saturation or

Temperature measurement mode.

2) With the MODE key pressed, power the unit on by pressing the ON key.

Release the ON key before releasing the MODE key.

3) “SEt.P” indicator will appear momentarily and “COF.1” will appear next.

4) You are now in the % Saturation or Temperature Set Up menu.

7.1.2

To enter the mg/L (ppm) Concentration Set Up menu:

5) Switch off the meter in the mg/L (ppm) Concentration measurement mode.

6) Repeat steps 2 to 3 as above and you will be in the mg/L (ppm)

Concentration Set Up menu.

The following table combines the explanation for the features available in both
Set Up menus. Figures 19

and 20 give an overall view of the Set Up sequence

for the % Saturation or Temperature Set Up menu and the mg/L (ppm)
Concentration Set Up menu.