7.3.3 Procedure for calibrating instrument with masters by external input

(1) Description

Set the minimum and maximum masters in that order.

(2) Mastering

Check that “READY” is displayed.

Set the minimum master in the measuring head.

After about 2 sec of stabilization time, input the “Minimum mastering” signal.

Set the maximum master in the measuring head.

After about 2 sec of stabilization time, input the “Maximum mastering” signal.

Check that “MAS OK” is displayed. If “MAS OK” is displayed, the mastering

procedure is completed. If “MAS OK” is not displayed, check the setting of each
master, and adjust the measuring head and A/E converter, among other things.

(3) Program change
When the program change on condition is held for 0.2 sec, the measuring program is
changed. Programs 1 to 3 can be freely changed from one to another.
*No change can be made to program 4.