KYORITSU 5020 Instruction Manual User Manual

Page 26

background image

̶

23

̶

5. Detection level at CH1 is indicated.

Can be set at every 1 from 0 to 1000

 ◆

To change the setting, press the

button.

Then the indication on the LCD flashes.

 ◆

Press the

or

button to set the

value to the desired one. Then press the

button to confirm it.

 ◆

When using Leakage Current Sensor, pressing

the

button at 1000mA indication

changes to 2A, pressing the

button at

2A indication changes to 1000mA (1A)

Point

  

Further detection doesn't start when a value falls

to 50% or less (for voltage sensor: 1% or more)

of the detected level. Set the appropriate value

based on the test prior to recording.

6. Confirm or change the preset current value on

channel 2 and 3 as well.

7. Next, One-time/ Endless is indicated.

One-time: Recording stops when memory

becomes full.

Endless : Overwrite the old data, and store

the latest data.

 ◆

Press the

button and proceed to the

next step when not changing the setting.

 ◆

To change the setting, press the

button.

Then the indication on the LCD flashes.

  

Press the

or

button to set

the value to the desired one. Then press the

button to confirm it.