KYORITSU 5020 Instruction Manual User Manual

Page 15

background image

̶

12

̶

4. Can make measurement right after

powering on the instrument.

* Each time pressing the

button,

CH1 to CH3 switches. When

(non-connect) is displayed on the
LCD, a sensor is not connected
to the selected channel; or the
connection is incorrect.

Check the mark indicating the selected
recording mode. Refer to “5. Recording
modes and conditions
” in this manual
to change the recording mode or
condition (Recording interval / Preset
current)

Recording mode

Details

Refer to

Normal

Performs continuous measurement/
recording at the pre-set intervals.
(15-kind: 1 sec. to 60min.)

P.19

Trigger

Records 8 data (0.8 sec.) in total with
time information when pre-set detection
level is exceeded.

P.22

Capture

Records 10 (50Hz) to 12 waveforms
(60Hz) when preset detection level is
exceeded.
(Graph can be viewed on PC)

P.24

(PC setting)

Power Quality Analysis

(only 5020)

Records date and time information when
preset Swell, Dip and Short interruption
value are exceeded.

P.24

(PC setting)

Step2: Confirmation of preset value