beautypg.com

KYORITSU 5020 Instruction Manual User Manual

Page 18

background image

̶

15

̶

* Settings cannot be changed during recording. Carefully check the
settings prior to a recording.
* Previous recorded data is cleared and new recording starts when
changing the followings and starts recording. (In this case, "CLr" flashes on
the LCD at a start of recording.)
1) Recording mode is changed.
2) CH for Sensor is changed.
3) Sensor type is changed.
* Transfer important data to your PC first and clear Memory (P.33) since
further recording isn't activated when all the 3 blocks in Memory Block
(P.15) are full.
* Follow the procedure below to start recording.

1. Press down the

button at least 4 sec.

"

" or "CLr" mark flashes. Keep the button

pressed down until "

" mark lights up.

Recording stops when the button is released
while the mark is flashing.

2. Measured value is displayed and "

" mark

lights up, then measurement starts.

Additional three recordings (Memory Block /
see the next page) are possible.

In this case, one-time is automatically selected
and recording stops when memory becomes
full.

It is recommended to clear the memory before
starting to record if previous data is not
necessary.

Step3: Preparation before recording

or

This manual is related to the following products: