KROHNE ALTOSONIC V Modbus EN User Manual
Page 47
ALTOSONIC V
ModBus Manual 0300 rev07 E
7.30855.35.00
Page 47 of 64
30
59
F32+RW
UFP batch control:
Normal:
Setup=9 (if UFP batch1 status batch=0 is no batch)
Cancel=5 (if UFP batch1 status batch=1 is set-up)
Start batch=119 (if UFP batch1 status batch=1 is set-up)
End batch=229 (if UFP batch1 status batch=1 is running)
reset print=1009 (if UFP batch1 status batch=5…10 printing)
Confirm ticket=779 (if UFP batch1 status batch=10 is confirm)
Continuous pipe line measurement, ticket on demand:
End with no reset values=559 (if UFP batch1 status batch is not
printing)
End with reset values=229 (if UFP batch1 status batch is not
printing)
Reset printing=1009 (if UFP batch1 status batch 5…10
is printing)
559 and 229 have automatic freeze of batch values (on Modbus)
for a maximum of 30 seconds (see also Boolean 2075)
6
2
5…1009
On command value input:
Return 0 if not accepted
Return 1 if accepted
Reset to -99999 after 5 seconds
31
61
F32+R
Solartron1: time to update a parameter (read only)
5
2
s, max 30 sec
32
63
F32+RW
Solartron1 K0
6
2
device calib. Param.
33
65
F32+RW
Solartron1 K1
6
2
device calib. Param.
34
67
F32+RW
Solartron1 K2
6
2
device calib. Param.
35
69
F32+RW
Solartron1 K18
6
2
device calib. Param.
36
71
F32+RW
Solartron1 K19
6
2
device calib. Param.
37
73
F32+RW
Solartron1 K20A
6
2
device calib. Param.
38
75
F32+RW
Solartron1 K20B
6
2
device calib. Param.
39
77
F32+RW
Solartron1 K21A
6
2
device calib. Param.
40
79
F32+RW
Solartron1 K21B
6
2
device calib. Param.
41
81
F32+R
Solartron2: time to update a parameter (read only)
5
2
s, max 30 sec
42
83
F32+RW
Solartron2 K0
6
2
device calib. Param.
43
85
F32+RW
Solartron2 K1
6
2
device calib. Param.
44
87
F32+RW
Solartron2 K2
6
2
device calib. Param.
45
89
F32+RW
Solartron2 K18
6
2
device calib. Param.
46
91
F32+RW
Solartron2 K19
6
2
device calib. Param.
47
93
F32+RW
Solartron2 K20A
6
2
device calib. Param.
48
95
F32+RW
Solartron2 K20B
6
2
device calib. Param.
49
97
F32+RW
Solartron2 K21A
6
2
device calib. Param.
50
99
F32+RW
Solartron2 K21B
6
2
device calib. Param.
51
101
F32+R
Sarasota1: time to update a parameter (read only)
5
2
s, max 30 sec
52
103
F32+RW
Sarasota1 K
6
2
device calib. Param.
53
105
F32+RW
Sarasota1 T0
6
2
device calib. Param.
54
107
F32+RW
Sarasota1 D0
6
2
device calib. Param.
55
109
F32+RW
Sarasota1 Nt
6
2
device calib. Param.
56
111
F32+RW
Sarasota1 Np
6
2
device calib. Param.
57
113
F32+RW
Sarasota1 Tcal
6
2
device calib. Param.
58
115
F32+RW
Sarasota1 Pcal
6
2
device calib. Param.
59
117
F32+R
Sarasota2: time to update a parameter (read only)
5
2
s, max 30 sec
60
119
F32+RW
Sarasota2 K
6
2
device calib. Param.
61
121
F32+RW
Sarasota2 T0
6
2
device calib. Param.
62
123
F32+RW
Sarasota2 D0
6
2
device calib. Param.
63
125
F32+RW
Sarasota2 Nt
6
2
device calib. Param.
64
127
F32+RW
Sarasota2 Np
6
2
device calib. Param.
65
129
F32+RW
Sarasota2 Tcal
6
2
device calib. Param.
66
131
F32+RW
Sarasota2 Pcal
6
2
device calib. Param.
67
133
F32+RW
Input in UFP (if enabled): temperature body
6
1
°C
68
135
F32+RW
Input in UFP (if enabled): temperature process
6
2
°C
69
137
F32+RW
Input in UFP (if enabled): temperature proving. external flowmet
6
8
°C
70
139
F32+RW
Input in UFP (if enabled): temperature densitometer
6
2
°C
71
141
F32+RW
Input in UFP (if enabled): pressure process
6
2
bar