ALTOSONIC V

ModBus Manual 0300 rev07 E

7.30855.35.00

Page 47 of 64

30

59

F32+RW

UFP batch control:
Normal:
Setup=9 (if UFP batch1 status batch=0 is no batch)
Cancel=5 (if UFP batch1 status batch=1 is set-up)
Start batch=119 (if UFP batch1 status batch=1 is set-up)
End batch=229 (if UFP batch1 status batch=1 is running)
reset print=1009 (if UFP batch1 status batch=5…10 printing)
Confirm ticket=779 (if UFP batch1 status batch=10 is confirm)
Continuous pipe line measurement, ticket on demand:
End with no reset values=559 (if UFP batch1 status batch is not
printing)
End with reset values=229 (if UFP batch1 status batch is not
printing)
Reset printing=1009 (if UFP batch1 status batch 5…10
is printing)
559 and 229 have automatic freeze of batch values (on Modbus)
for a maximum of 30 seconds (see also Boolean 2075)

6

2

5…1009
On command value input:
Return 0 if not accepted

Return 1 if accepted
Reset to -99999 after 5 seconds

31

61

F32+R

Solartron1: time to update a parameter (read only)

5

2

s, max 30 sec

32

63

F32+RW

Solartron1 K0

6

2

device calib. Param.

33

65

F32+RW

Solartron1 K1

6

2

device calib. Param.

34

67

F32+RW

Solartron1 K2

6

2

device calib. Param.

35

69

F32+RW

Solartron1 K18

6

2

device calib. Param.

36

71

F32+RW

Solartron1 K19

6

2

device calib. Param.

37

73

F32+RW

Solartron1 K20A

6

2

device calib. Param.

38

75

F32+RW

Solartron1 K20B

6

2

device calib. Param.

39

77

F32+RW

Solartron1 K21A

6

2

device calib. Param.

40

79

F32+RW

Solartron1 K21B

6

2

device calib. Param.

41

81

F32+R

Solartron2: time to update a parameter (read only)

5

2

s, max 30 sec

42

83

F32+RW

Solartron2 K0

6

2

device calib. Param.

43

85

F32+RW

Solartron2 K1

6

2

device calib. Param.

44

87

F32+RW

Solartron2 K2

6

2

device calib. Param.

45

89

F32+RW

Solartron2 K18

6

2

device calib. Param.

46

91

F32+RW

Solartron2 K19

6

2

device calib. Param.

47

93

F32+RW

Solartron2 K20A

6

2

device calib. Param.

48

95

F32+RW

Solartron2 K20B

6

2

device calib. Param.

49

97

F32+RW

Solartron2 K21A

6

2

device calib. Param.

50

99

F32+RW

Solartron2 K21B

6

2

device calib. Param.

51

101

F32+R

Sarasota1: time to update a parameter (read only)

5

2

s, max 30 sec

52

103

F32+RW

Sarasota1 K

6

2

device calib. Param.

53

105

F32+RW

Sarasota1 T0

6

2

device calib. Param.

54

107

F32+RW

Sarasota1 D0

6

2

device calib. Param.

55

109

F32+RW

Sarasota1 Nt

6

2

device calib. Param.

56

111

F32+RW

Sarasota1 Np

6

2

device calib. Param.

57

113

F32+RW

Sarasota1 Tcal

6

2

device calib. Param.

58

115

F32+RW

Sarasota1 Pcal

6

2

device calib. Param.

59

117

F32+R

Sarasota2: time to update a parameter (read only)

5

2

s, max 30 sec

60

119

F32+RW

Sarasota2 K

6

2

device calib. Param.

61

121

F32+RW

Sarasota2 T0

6

2

device calib. Param.

62

123

F32+RW

Sarasota2 D0

6

2

device calib. Param.

63

125

F32+RW

Sarasota2 Nt

6

2

device calib. Param.

64

127

F32+RW

Sarasota2 Np

6

2

device calib. Param.

65

129

F32+RW

Sarasota2 Tcal

6

2

device calib. Param.

66

131

F32+RW

Sarasota2 Pcal

6

2

device calib. Param.

67

133

F32+RW

Input in UFP (if enabled): temperature body

6

1

°C

68

135

F32+RW

Input in UFP (if enabled): temperature process

6

2

°C

69

137

F32+RW

Input in UFP (if enabled): temperature proving. external flowmet

6

8

°C

70

139

F32+RW

Input in UFP (if enabled): temperature densitometer

6

2

°C

71

141

F32+RW

Input in UFP (if enabled): pressure process

6

2

bar