ALTOSONIC V

ModBus Manual 0300 rev07 E

7.30855.35.00

Page 23 of 64

7.3

Software set-up

Now set-up the software, all the settings for the Modbus driver is done in the file [coms0300.dat].

See also chapter 9.4 Appendix D: Coms0300.dat file

7.3.1

First set the parameters for the communication line

3.1 MODBUS_UART_BASEADRESS for channel 1 is COM4 this is baseaddress

0x2E8

3.2 MODBUS_UART_INTERRUPT is for COM4 set to interrupt

3.

Depends on your application : 3.3 MODBUS_UART_BAUDRATE 1200,2400,4800,9600,19200

3.4 MODBUS_UART_RTS_MODE to

0.

Depends on your application : 3.5 MODBUS_UART_N_DATABITS to 7 or 8

Depends on your application : 3.6 MODBUS_UART_N_STOPBITS to 1 or 2

Depends on your application : 3.7 MODBUS_UART_PARITY to none, even or odd.

Depends on your application : 3.3 MODBUS_UART_BAUDRATE 1200,2400,4800,9600,19200

Depends on your application :
If you use RS485 set 3.8 MODBUS_UART_HALF_DUPLEX to HALF_DUPLEX(=1)
If you use RS422 set 3.8 MODBUS_UART_HALF_DUPLEX to FULL_DUPLEX(=0)

7.3.2

Now select the parameters for the used protocol

Select the frame type RTU or ASCII with 3.9 MODBUS_TRANSFER_MODE.

Set the UFP-V as MASTER or SLAVE device with 5.1 MODBUS_DEVICE_TYPE.

Select if variables, which are larger than 16 bits are still counted as the number of 16 bit

Set the data points requesting type by parameter 5.2 MODBUS_MODICON_COMPAT:
By type is not modicon compitable ( =0)
By 16 bit registers is modicon compitable ( =1)

7.3.3

The UFP-V as SLAVE device

The slave mode is activated when the parameter 5.1 MODBUS_DEVICE_TYPE=1.

If the UFP-V acts like a Modbus Slave device, set the SlaveID with 5.3 MODBUS_SLAVE_ID.

The 5.4 FLAG_HOLD_TIME is a hold time on the status flags (Booleans only).
The 5.4 FLAG_HOLD_TIME freezes the flags after the flag has changed from state.

Set this time a bit larger than the maximum communication-request interval.

The next fields define to which Modbus addresses the data of the UFP-V is mapped to, these

settings are default settings and should not be changed, only if necessary.
The fields are 6 DATAFIELD 1 to N, for every DATAFIELD an access mode could be set.
The 6 ACCES MODE defines how the data is send and interpreted when the UFP-V is in slave-
mode
.

See the manual of the accompanying byte-order of transmission/reception with the 2 modes.

For Slave-use the driver should be working now.