Profibus settings, 4 menu d, service – KROHNE MFC 400 Profibus User Manual
Page 51
PROFIBUS SETTINGS
5
51
MFC 400
www.krohne.com
06/2013 - 4002835301 - AD MFC 400 PROFIBUS R01 en
5.4 Menu D, Service
This menu is protected. You will need to use the service password to gain access.
C6.8.1
Station Address
Selects the PROFIBUS station address of the device.
The PROFIBUS address can also be changed using the PROFIBUS service
"set_slave_add". The input range is 0...125 according to the PROFIBUS
specification. Address 126 is the default address and cannot be set via the
PROFIBUS service "set_slave_add" - use menu instead to reset to default
address.
C6.8.2
Information
Available if there is a PROFIBUS interface in existence; the below mentioned
information will be displayed about the PROFIBUS interface:
Hardware Revision of the PROFIBUS interface used
Software Version of the PROFIBUS interface used
Date of Compilation
(same as B3.5)
C6.8.3
Diag. extension
Content of the PROFIBUS diagnosis extension is displayed.
C6.8.4
Diag. extension i
Content of the internal diagnosis extension is displayed.
C6.8.5
Diag. extension h
h = history
Content of the PROFIBUS diagnosis extension history is displayed. Shows
PROFIBUS diagnosis extension information of all PROFIBUS diagnosis
extension bits which are set since last reset (all bits reset to zero – use
service menu D2.2.9 to reset this parameter)
C6.8.6
Diag. extension ih
h = history
Content of the internal PROFIBUS diagnosis extension history is displayed.
Shows internal PROFIBUS diagnosis extension information of all internal
PROFIBUS diagnosis extension bits which are set since last reset (all bits
reset to zero – use service menu D2.2.10 to reset this parameter)
C6.8.7
Identification No.
Sets different device modes for the cyclic configuration of the PROFIBUS
interface.
Select:
MFC400: all manufacturer specific parameters / functions supported
Profile: profile defined parameters / functions supported only
Autom.: (default setting) PROFIBUS configuration (MFC400 or Profile) will be
selected automatically by the PROFIBUS master itself without interaction by
the user. The active selection is configured during startup by using the
Set_Prm and/or Set_Slave_Add service. No additional acyclic parameter
settings required.
C6.8.8
Condensed Status
Defines the general method how the device handles status and diagnostics:
On: Condensed status and diagnosis information according PROFIBUS
Profile 3.02 will be used
Off: Classic status and diagnosis information according PROFIBUS Profile
3.02 will be used
No.
Function
Settings / descriptions
D2.2
Service Parameter
D2.2.7
Identification No.
Same as C6.8.7
D2.2.8
PB Cold Start
Initiates a PROFIBUS cold start.
During a PROFIBUS cold start nearly all parameter values of the whole
device will be set to their default values (Exceptions are the PROFIBUS
address and the identification No.). The PROFIBUS cold start will be carried
out without disconnecting an already established connection to a PROFIBUS
master system.
Select:
No (terminates this function) / Yes (PROFIBUS cold start will be carried out
at once)
D2.2.9
Diag. extension h
Reset (to zero) the whole contents of the Diag. extension h parameter
D2.2.10
Diag. extension ih
Reset (to zero) the whole contents of the Diag. extension ih parameter