Detailed description of special settings concerning the PROFIBUS features easily operated via
the local display menu of the signal converter (refer to the following sections). For a detailed
description of the menus and functions in general please refer to the standard product
documentation of the signal converter.
5.1 Menu A, Quick Setup
5.2 Menu B, Test
No.
Function
Setting / Description
A2 Tag
A2
Tag
Tag descriptor of the physical block of the PROFIBUS module will be
displayed. The Tag descriptor provides an application specific reference to
the blocks. It will be assigned by the user of the device. The Tag descriptor is
an octet string (a visible string will be preferred) consisting of 32 byte. The
first 21 characters will be displayed only!
A3 Reset
A3
Reset
A3.1
Reset Errors
Reset Errors? Select: no/yes
The following reset menus for the Totalisers are available only if the "quick access" has been activated in the menu
"Setup → Device → Quick Setup (C6.9.x)". Each Totaliser can be activated independently for quick access.
A3.2
All Totalisers
All Totalisers can be reset to zero at the same time.
Reset all Totaliser? Select: No / Yes (available if activated in C6.9.4)
A3.3
FB4 Totaliser 1
Reset Totaliser 1 only? Select: No / Yes (available if activated in C6.9.1)
A3.4
FB5 Totaliser 2
Reset Totaliser 2 only? Select: No / Yes (available if activated in C6.9.2)
A3.5
FB6 Totaliser 3
Reset Totaliser 3 only? Select: No / Yes (available if activated in C6.9.3)
A4 Station Address
A4
Station Address
Selects the address of the device at the PROFIBUS interface.
The PROFIBUS address can also be changed using the PROFIBUS service
"set_slave_add". The input range is 0...125 according to the PROFIBUS
specification. Address 126 is the default address and cannot be set via the
PROFIBUS service "set_slave_add" - use menu instead to reset to default
address.
No.
Function
Settings / descriptions
B3 Information
B3.5
PROFIBUS
Available if there is a PROFIBUS interface in existence; the below mentioned
information will be displayed about the PROFIBUS interface:
Hardware Revision of the PROFIBUS interface used
Software Version of the PROFIBUS interface used
Date of Compilation
(same as C8.2)