PROFIBUS SETTINGS
5
49
MFC 400
www.krohne.com
06/2013 - 4002835301 - AD MFC 400 PROFIBUS R01 en
C5 I/O PROFIBUS
C5
I/O PROFIBUS
Using the menu functions mentioned below you will be able to control
basically the five analog input blocks of this PROFIBUS device. These eights
menus are identical so they are grouped together and their functions are
described in one go.
C5.1
FB1 Analog Inp.
There are 8 analog input blocks.
_ stands for the 8 analog input blocks: FB1 (_ = 1), FB2 (_ = 2), FB3 (_ = 3),
FB7 (_ = 4), FB8 (_ = 5), FB9 (_ = 6), FB10 (_ = 7), FB11 (_ = 8),
C5.2
FB2 Analog Inp.
C5.3
FB3 Analog Inp.
C5.4
FB7 Analog Inp.
C5.5
FB8 Analog Inp.
C5.6
FB9 Analog Inp.
C5.7
FB10 Analog Inp.
C5.8
FB11 Analog Inp.
C5._.1
Measurement
Select measurement value for the analog input blocks:
Volume Flow / Mass Flow / Density / Temperature / Concentration 1 /
Flow Velocity / Electronics Temp. / Supply / Concentration 2 /
Conc. volume flow 1 / Conc. mass flow 1 / Conc. volume flow 2 /
Conc. mass flow 2 / Sensor Average / Sensor Deviation / Drive Level /
Tube Frequency / Strain 1 / Strain 2 / 2 Phase Signal
C5._.2
Time Constant
Set time constant for this function block (_).
C5._.3
Error Behaviour
Defines the behaviour of this function block in case of errors.
Select: hold value / ignore error / replace value
hold value:
hold value:
hold value:
hold value: Last valid OUT value stored will be used as OUT value.
ignore error:
ignore error:
ignore error:
ignore error: OUT has the wrong calculated value and status "Bad" as
calculated.
replace value:
replace value:
replace value:
replace value: The "replacement value" will be used as OUT value.
C5._.4
Replacement Value
Available, if the error behaviour "Replace Value" is selected. Defines the
value that replaces the measured value at this function block in case of an
error.