COMMISSIONING / OPERATION

4

33

MFC 400

www.krohne.com

06/2013 - 4002835301 - AD MFC 400 PROFIBUS R01 en

63

3

F – Sensor

(classification fixed - not changeable)

Sensor is defective and not able to measure. Related output

values are invalid. Typically a repair or replacement of the device

is required. Check sensor / cabling / 4 - restart the device – if the

error still occur send back the device to the manufacturer with an

indication of the error.

64

3

M – Process

(classification fixed - not changeable)

Process related problem(s) detected. But still the measurement

values are accurate. The output values are still valid, but this may

change in the near future. Check problem(s) / 5 - if the

problem(s) cannot be fixed contact the service department of the

manufacturer with an indication of the problem(s).

65

3

M – Configuration

(classification fixed - not changeable)

Configuration related problem(s) detected. But still the

measurement values are accurate. The output values are still

valid, but this may change in the near future. Check problem(s) /

5 - if the problem(s) cannot be fixed contact the service

department of the manufacturer with an indication of the

problem(s).

66

3

M – Electronics

(classification fixed - not changeable)

Converter Electronic related problem(s) detected. But still the

measurement values are accurate. The output values are still

valid, but this may change in the near future. Check problem(s) /

5 - if the problem(s) cannot be fixed contact the service

department of the manufacturer with an indication of the

problem(s).

67

3

F – Sensor

(classification fixed - not changeable)

Sensor related problem(s) detected. But still the measurement

values are accurate. The output values are still valid, but this may

change in the near future. Check problem(s) / 5 - if the

problem(s) cannot be fixed contact the service department of the

manufacturer with an indication of the problem(s).

68

3

S – Process

(classification fixed - not changeable)

Process related fault(s) detected. E.g. the device may be operated

outside its specified operating range. The measurement accuracy

may be reduced. But still the measurement value follows the

source. Check fault(s) detected / 4 - if the fault(s) cannot be

cleared contact the service department of the manufacturer with

an indication of the problem(s).

69

3

S – Configuration

(classification fixed - not changeable)

Configuration related fault(s) detected. E.g. the device may be

operated outside its specified operating range. The measurement

accuracy may be reduced. But still the measurement value

follows the source. Check fault(s) detected / 4 - if the fault(s)

cannot be cleared contact the service department of the

manufacturer with an indication of the problem(s).

70

3

S – Electronics

(classification fixed - not changeable)

Converter Electronic related fault(s) detected. E.g. the device may

be operated outside its specified operating range. The

measurement accuracy may be reduced. But still the

measurement value follows the source. Check fault(s) detected /

4 - if the fault(s) cannot be cleared contact the service

department of the manufacturer with an indication of the

problem(s).

71

3

S – Sensor

(classification fixed - not changeable)

Sensor related fault(s) detected. E.g. the device may be operated

outside its specified operating range. The measurement accuracy

may be reduced. But still the measurement value follows the

source. Check fault(s) detected / 4 - if the fault(s) cannot be

cleared contact the service department of the manufacturer with

an indication of the problem(s).

72

3

("I-") Electr: Power Failure

(default classification / 9)

Information only; e.g. power-down detected / 7. Just information.

No impact on measurement expected.

73

3

("I-") Proc: System Control

(default classification / 9)

Information only; e.g. system-control messages present / 7. Just

information. No impact on measurement expected.

Bit Number

sup

p

orted

De

sc

ription

Remedy