Operation – KROHNE OPTISYS SLM 2100 EN User Manual
Page 56

5
OPERATION
56
OPTISYS SLM 2100
www.krohne.com
03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en
Step 3: calibrate zero
The calibration zero is done.
• Press ^
^
^
^ several time until you reach the measuring mode again. Choose yes
yes
yes
yes to safe and
confirm your selection.
Step 3a: calibrate conc. with known reference value
Press
or
until the submenu calibrate zero
calibrate zero
calibrate zero
calibrate zero is highlighted.
Press >>>> to enter the chosen menu.
actual value
actual value
actual value
actual value is shown.
Press ^
^
^
^ to enter the chosen menu.
calibrate zero
calibrate zero
calibrate zero
calibrate zero is shown.
Press
or
to select calibrate
calibrate
calibrate
calibrate to calibrate or break
break
break
break to interrupt the calibration.
Press ^
^
^
^ to enter the chosen menu.
Start the calibration process with step 2: move sensor
Press
or
until the submenu calibrate conc.
calibrate conc.
calibrate conc.
calibrate conc. is highlighted.
Press >>>> to enter the chosen menu.
reference value
reference value
reference value
reference value is shown.
Press
or
to enter the reference value.
Press ^
^
^
^ to confirm the value.
ADC value
ADC value
ADC value
ADC value is shown.
Press
or
to choose between stored value
stored value
stored value
stored value to use the stored value or actual value
actual value
actual value
actual value to use
the actual value for calibration.
Press
or
to select actual value
actual value
actual value
actual value.
Press ^
^
^
^ to enter the chosen menu.
actual value
actual value
actual value
actual value is shown.
Press ^
^
^
^ to confirm the value.
use value
use value
use value
use value is shown.
Press
or
to choose between break
break
break
break to interrupt the calibration,calibrate
calibrate
calibrate
calibrate
to adjust the actual value to the reference value or save ADC value
save ADC value
save ADC value
save ADC value to save
the actual value.
Press
or
to select calibrate
calibrate
calibrate
calibrate
Press ^
^
^
^ to enter the chosen menu.
Press ^
^
^
^ several time until you reach the measuring mode again.
Choose yes
yes
yes
yes to safe and confirm your selection.