beautypg.com

Electrical connections – KROHNE OPTISYS SLM 2100 EN User Manual

Page 30

background image

4

ELECTRICAL CONNECTIONS

30

OPTISYS SLM 2100

www.krohne.com

03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en

4.9.2 Current output (active)

The terminals for the connection of the two current outputs are located on the main board.
Please refer to the following diagram for proper connection of the cables.

DANGER!
All work on the electrical connections may only be carried out with the power disconnected.

CAUTION!
To avoid damage or destruction of the device always note the following items:

Observe the connection polarity!

Note the properties of the current output; for further information refer to Technical data on
page 68.

INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.

Connection diagram of current output (active)

Figure 4-4: Current output (active)

Figure 4-5: Connection current output

1 Current output A
2 Current output B

INFORMATION!
The correct connection of the current outputs takes place with the help of shielded signal cables
which are approved for the rated current and voltage.

.book Page 30 Wednesday, April 30, 2014 1:39 PM