B.84 SYSTem:FAN COMMAND
SYST:FAN
Syntax:
Short Form: SYST:FAN
Long Form: SYSTem:FAN
Where
Description: SYST:FAN 0 sets fan to off. SYST:FAN 100 sets fan to on.
B.85 SYSTem:FAN? QUERY
SYST:FAN?
Syntax:
Short Form: SYST:FAN?
Long Form: SYSTem:FAN?
Returns
Description: Indicates fan status: 0 = off, 100 = on.
B.86 0SYSTem:PFModule:DAMPing COMMAND
SYST:PFM:DAMP
Syntax:
Short Form: SYST:PFM:DAMP
Long Form: SYSTem:PFModule:DAMPing
Where
Description: Establishes damping settings for Power FET Modules. Increased damping decreases the loop's
response time to transients. Factory default setting is 3.
Damping Settings: 0 - 1nF, 1 - 4.3nF, 2 - 11nF, 3 - 34nF, 4 - 101nF, 5 - 147.3nF. See PAR. 3.14 for
details.
B.87 SYSTem:PFModule:DAMPing? QUERY
SYST:PFM:DAMP?
Syntax:
Short Form: SYST:PFM:DAMP?
Long Form: SYSTem:PFModule:DAMPing?
Returns
Description: Indicating damping settings for Power FET Modules. See PAR. 3.14 for details.
B.88 SYSTem:PFModule:LIST? QUERY
SYST:PFM:LIST?
Syntax:
Short Form: SYST:PFM:LIST?
Long Form: SYSTem:PFModule:LIST?
Description: List Power FET Modules within chassis (see below for example).
SYST:PFM:LIST?
KEPCO, EL 5K-200-600 01-18-2010,A999104,MCB #200 1.176 $ 2010/01/18 22:27:06 $
6 Power FET Modules Installed
Pos Serial
Revision
0 292
1.39
1 293
2 294
1.39
3 310
1.39
4 1200
1.39
5 296
NOTE: PFMs with no revision are using Firmware Revisions below 1.35.