B.42 [source:]voltage:protection:over command, B.43 [source:]voltage:protection:over? query, B.46 [source:]voltage:protection:under command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.63 through 3.86 User Manual
Page 69: B.40, Source, B.41, B.42, B.43, B.44, B.45

SERIES EL 070312
B-7
B.40 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude] COMMAND
VOLT
Syntax:
Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP]
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]
where:
Description: Establishes voltage setpoint. The voltage setpoint only affects the load when the load is in Voltage
mode of operation. Voltage setpoint is stored in non-volatile memory; value is restored upon power-up.
B.41 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]? QUERY
VOLT?
Syntax:
Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]?
Returns:
Description: Indicates voltage setpoint.
B.42 [SOURce:]VOLTage:PROTection:OVEr COMMAND
VOLT:PROT:OVE
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVE
Long Form: [SOURce:]VOLTage:PROTection:OVEr
where
Description: Sets overvoltage protection limit. Overvoltage protection setpoint is stored in non-volatile memory;
value is restored upon power-up.
B.43 [SOURce:]VOLTage:PROTection:OVEr? QUERY
VOLT:PROT:OVE?
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVE?
Long Form: [SOURce:]VOLTage:PROTection:OVEr?
Returns
Description: Indicates overvoltage protection limit.
B.44 [SOURce:]VOLTage:PROTection:OVEr:STATe[:LEVel] COMMAND
VOLT:PROT:OVE:STAT 0
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVE:STAT[:LEV] 0
Long Form: [SOURce:]VOLTage:PROTection:OVEr:STATe[:LEVel] 0
Description: Clears overvoltage protection error.
B.45 [SOURce:]VOLTage:PROTection:OVEr:STATe[:LEVel]? COMMAND
VOLT:PROT:OVE:STAT?
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVE:STAT[:LEV]
Long Form: [SOURce:]VOLTage:PROTection:OVEr:STATe[:LEVel]
where
0 (overvoltage protection error has not occurred)
Description: Indicates whether an overvoltage protection has occurred (1) or not (0).
B.46 [SOURce:]VOLTage:PROTection:UNDer COMMAND
VOLT:PROT:UND
Syntax:
Short Form: [SOUR:]VOLT:PROT:UND
Long Form: [SOURce:]VOLTage:PROTection:UNDer
where
Description: Sets undervoltage protection limit. Undervoltage protection setpoint is stored in non-volatile mem-
ory; value is restored upon power-up.