B.16 [SOURce:]OUTPut:PROTection:SHORt[:STATe] COMMAND

OUTP:PROT:SHOR

Syntax:

Short Form: [SOUR:]INP:PROT:SHOR[:STAT]
Long Form: [SOURce:]INPut:PROTection:SHORt[:STATe]

Description: Applies maximum drive to FETs, effectively shorting the output.

B.17 [SOURce:]MEASure:ALL? QUERY

MEAS:ALL?

Syntax:

Short Form: [SOUR:]MEAS:ALL? Long

Form:

[SOURce:]MEASure:ALL?

Returns: Measured values in the form 0.000 KW, 0.000 Amps, 0.000 Volts.

Description: Measures output power (Kilowatts), current (Amps), and voltage (Volts). Unit of measurement included

in the reply.

B.18 [SOURce:]MEASure:ALL2? QUERY

MEAS:ALL2?

Syntax:

Short Form: [SOUR:]MEAS:ALL2? Long

Form:

[SOURce:]MEASure:ALL2?

Returns: Measured values in the form 0.000 0.000 0.000 (Kilowatts Amps Volts).

Description: Measures output power (Kilowatts), current (Amps), and voltage (Volts). Unit of measurement is omit-

ted from the reply.

B.19 [SOURce:]MEASure[:SCALar]:CURRent[:DC]? QUERY

MEAS:CURR?

Syntax:

Short Form: [SOUR:]MEAS[:SCAL]:CURR[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:CURRent[:DC]?
Returns: Measured values.

Description: Measures current. Measured value returned as XXX Amps, where XXX is decimal data representing

measured current.

B.20 [SOURce:]MEASure[:SCALar]:POWer[:DC]? QUERY

MEAS:POW?

Syntax:

Short Form: [SOUR:]MEAS[:SCAL]:POW[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:POWer[:DC]?
Returns: Measured values in the form 0.000 Kilowatts

Description: Measures power. Measured value returned as XXX Kilowatts, where XXX represents measured

power.

B.21 [SOURce:]MEASure[:SCALar]:VOLTage[:DC]? QUERY

MEAS:VOLT?

Syntax:

Short Form: [SOUR:]MEAS[:SCAL]:VOLT[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:VOLTage[:DC]?
Returns: Measured values in the form XXX Volts

Description: Measures voltage. Measured value returned as XXX Volts, where XXX is decimal data representing

measured voltage.

B.22 [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude] COMMAND

COND

Syntax:

Short Form:[SOUR:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]
Long Form: [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]
Where = conductance setpoint in Siemens

Description: Establishes conductance setpoint. Conductance setpoint affects the load only when the load is in

Conductance mode of operation. Conductance setpoint is stored in non-volatile memory; value is
restored upon power-up.

B.23 [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]? QUERY

COND?

Syntax:

Short Form:[SOUR:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]?
Long Form: [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]?
Returns = conductance setpoint in Siemens

Description: Indicates conductance setpoint.