B.24 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND

CURR

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]
Where = current setpoint

Description: Current Setting. The current setpoint affects the load only when the load is in Current mode of opera-

tion. Current setpoint is stored in non-volatile memory; value is restored upon power-up.

B.25 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? QUERY

CURR?

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]?
Returns = current setpoint

Description: Indicates Current Setpoint.

B.26 [SOURce:]CURRent:PROTection[:LEVel] COMMAND

CURR:PROT

Syntax:

Short Form: [SOUR:]CURR:PROT[:LEV]
Long Form: [SOURce:]CURRent:PROTection[:LEVel]
where: = current protection setpoint

Description: Establishes current protection setpoint. Current protection setpoint is stored in non-volatile mem-

ory; value is restored upon power-up. This is a user-defined value which, if exceeded during opera-
tion, causes a FAULT condition, forcing the load to disengage.

B.27 [SOURce:]CURRent:PROTection[:LEVel]? QUERY

CURR:PROT?

Syntax:

Short Form: [SOUR:]CURR:PROT[:LEV]?
Long Form: [SOURce:]CURRent:PROTection[:LEVel]?
Returns: = current protection setpoint

Description: Indicates current protection setpoint.

B.28 [SOURce:]CURRent:PROTection:STATe COMMAND

CURR:PROT:STAT 0

Syntax:

Short Form: [SOUR:]CURR:PROT:STAT 0
Long Form: [SOURce:]CURRent:PROTection:STATe 0

Description: Clears current protection level error.

B.29 [SOURce:]CURRent:PROTection:STATe? QUERY

CURR:PROT:STAT?

Syntax:

Short Form: [SOUR:]CURR:PROT:STAT?
Long Form: [SOURce:]CURRent:PROTection:STATe?
Returns: = 1 (current protection error occurred)

0 (current protection error has not occurred)

Description: Indicates whether a current protection error has occurred (1) or not (0).

B.30 [SOURce:]MODE COMMAND

MODE

Syntax:

Short Form: [SOUR:]MODE

Long Form: [SOURce:]MODE

where: = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.

Description: Sets operating mode: When the operating mode is changed, the input is set to off. Send INPut ON to

apply the new mode setting to the load.

B.31 [SOURce:]MODE? QUERY

MODE?

Syntax:

Short Form: [SOUR:]MODE?

Long Form: [SOURce:]MODE?

Returns: = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.

Description: Indicates operating mode.