Save/recall command – GW Instek GDS-1000 Series Programming Manual User Manual
Page 20
COMMAND
DETAILS
39
Note
Before using this command, select the
measurement channel. See the example below.
Example :measure:source
1
:measure:vmin?
Selects Channel 1, and
then measures the
minimum amplitude.
60B
:MEASure:VPP
Query
Description
Returns the peak-to-peak amplitude (difference
between maximum and minimum amplitude)
Same as: Measure key → F1~F5 → F3 (Vpp)
Syntax
< Long >
< Short >
:measure:vpp?
:meas:vpp?
Returns
Note
Before using this command, select the
measurement channel. See the example below.
Example :measure:source
1
:measure:vpp?
Selects Channel 1, and
then measures the
peak-to-peak
amplitude.
61B
:MEASure:VRMS
Query
Description
Returns the root-mean-square voltage.
Same as: Measure key → F1~F5 → F3 (Vrms)
Syntax
< Long >
< Short >
:measure:vrms?
:meas:vrms?
Returns
Note
Before using this command, select the
measurement channel. See the example below.
Example :measure:source
1
:measure:vrms?
Selects Channel 1, and
then measures the root
mean square voltage.
GDS-1000 Programming Manual
40
12B
Save/Recall Command
:MEMory
244H
40
:MEMory
245H
40
:MEMory
246H
41
:MEMory
247H
41
*RCL.............................................................................
248H
42
:REF
249H
42
:REF
250H
43
:REF
251H
43
*SAV.............................................................................
252H
44
62B
:MEMory
Set
Description
Recalls a panel setting from the internal memory.
Same as: Save/Recall key (recall) → F3
Syntax
< Long >
< Short >
:memory
:mem
Parameter
Internal memory
1 ~ 15
S1 ~ S15
Example :memory1:recall:setup
Recalls the settings from
the internal memory S1
63B
:MEMory
Set
Description
Recalls a waveform from the internal memory and
saves it to a reference waveform.
Same as: Save/Recall key (recall) → F4
Syntax
< Long >
< Short >
:memory
:mem
Parameter
Internal memory