beautypg.com

GW Instek GDS-1000 Series Programming Manual User Manual

Page 11

background image

COMMAND

DETAILS

21

:channel:invert

:channel:invert?

:chan:inv

:chan:inv?

Parameter

Channel

Channel invert

1/2 CH1/2

0 off

1 on

Example :channel1:invert

1

Inverts Channel 1

28B

:CHANnel:MATH

Set

Query

Description

Selects or returns the math operation type.
Same as: Math key → F1

Syntax

< Long >

< Short >

:channel:math

:channel:math?

:chan:math
:chan:math?

Parameter

Channel

Math operation

1/2

CH1 or CH2

0

Math off

1

Add

2

Subtract

3

FFT

Example1 :channel1:math

2

Channel 1 – Channel 2

Example2 :channel2:math

2

Channel 1 – Channel 2

Example3 :channel2:math

2

Runs FFT on Channel 2

29B

:CHANnel:OFFSet

Set

Query

Description

Sets or returns the offset level for a channel. The
offset level range depends on the vertical scale.

GDS-1000 Programming Manual

22

Syntax

< Long >

< Short >

:channel:offset

:channel:offset?

:chan:offs
:chan:offs?

Parameter

Channel Offset level

1/2 CH1/2 ±0.5 –0.5V

~

+0.5V

(2mV/div~50mV/div)

±5.0 –5.0V

~

+5.0V

(100mV/div~500mV/div)

±50.0 –50.0V ~ +50.0V

(1V/div ~ 5V/div)

Example :channel1:scale

1.00e–2

:channel1:offset 2.00e–2

Sets the Channel 1
scale to 10mV/div
Sets the Channel 1
offset to 20mV

30B

:CHANnel:PROBe

Set

Query

Description

Sets or returns the probe attenuation factor.
Same as: Channel key → F4

Syntax

< Long >

< Short >

:channel:probe

:channel:probe?

:chan:prob
:chan:prob?

Parameter

Channel Probe attenuation factor

1/2 CH1/2 0 1x

1 10x

2 100x

Example :channel1:probe

1

Sets the Channel 1
probe attenuation
factor to 10x