Voltmeter

32V full scale, 5 digits 0.4" LED display

Program
Accuracy

Voltage: ± (0.03% of reading + 10mV)
Current: ± (0.3% of reading + 10mA)

Read back
Accuracy

Voltage: ± (0.03% of reading + 10mV)
Current: ± (0.3% of reading + 10mA)

CH3 of 3303S

Voltage
Current

2.5V/3.3V/5.0V, ±5%
3A

Line

≤ 3mV

Load

≤ 5mV

Ripple & Noise ≤ 1mVrms (5Hz ~ 1MHz)

Insulation Chassis

and

Terminal

20MΩ or above (DC 500V)

Chassis and
AC cord

30MΩ or above (DC 500V)

Operation
Environment

Indoor use, Altitude: ≤ 2000m
Ambient temperature: 0 ~ 40°C
Relative humidity: ≤ 80%
Installation category: II, Pollution degree: 2

Storage
Environment

Ambient temperature: –10 ~ 70°C
Relative humidity: ≤ 70%

Power Source

AC 100V/120V/220V/230V±10%, 50/60Hz

Accessories

User manual x1
Test lead GTL-104A x 2, GTL-105A x 1
(Europe) Test lead GTL-203A x 1, GTL-204A x 2

Dimensions

210 (W) x 130 (H) x 265 (D) mm

Weight

Approx. 7kg

Options

USB cable

GTL-246

USB 2.0, A-B type

56