beautypg.com

Rcl

Page 51: Local

background image

REMOTE

CONTROL

Return parameter

GW INSTEK,GPD-X3303,SN: xxxxxxxx, Vx.xx
(Manufacturer, model name, serial number,
firmware version)

RCL

Description

Recalls a panel setting.

1 – 4: Memory 1 to 4

Response time

Minimum 10ms

Example RCL1

Recalls the panel setting stored
in memory 1

SAV

Description

Stores the panel setting.

1 – 4: Memory 1 to 4

Response time

Minimum 10ms

Example SAV1

Stores the panel setting in
memory 1

BAUD

Description

Sets the baud rate to 9600bps/57600bps
/115200bps.

0: 115200bps, 1: 57600bps, 2: 9600bps

Response time

Minimum 10ms

Example BAUD0

Sets the baud rate to 115200bps.

LOCAL

Description

Exits remote mode and sets the instrument to local
mode.

51