GW Instek PSM-Series Programming Manual User Manual
Page 32
PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
⎯ 61 ⎯
Figure 8: Status registers and related commands
The CONDition register is a read-only register which monitors the present
state of the instrument. The CONDition register updates in real time and the
inputs are not latched or buffered. When a condition monitored by the
CONDition register becomes true, the bit for that condition also becomes
true (1). When the condition is false, the bit is 0.
The read-only EVENt register latches any false-to-true change in condition.
Once the bit in the EVENt register is set, it is no longer affected by changes
in the corresponding bit of the CONDition register. The bit remains set until
the controller reads it. The command *CLS (Clear Status) clears the EVENt
register.
QUEStionable Status Registers.
Table 4 shows the bit designations of the 16 bit QUEStionable Status
Register.
Enable
Register
Event
Register
Condition
Register
To SBR
PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
⎯ 62 ⎯
Table 4: QUEStionable Status Register
Bit 15
Bit 14
Bit 13
Bit 12
Bit 11
Bit 10
Bit 9
Bit 8
∗
NU
NU
NU
NU
NU
Summary
OVP
NU
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
NU
NU
NU
NU
NU
NU
Summar
y Current
Summary
Voltage
The command STATus:QUEStionable:CONDtion? Reads the
QUEStionable CONDition register but dose not clear it.
The command STATus:QUEStionable:EVENt? Reads the
QUEStionable EVENt Status register and clears it.
OPERation Status Registers
Table 5 shows the bit designations of the 16 bit OPERation Status Register.
Table 5: OPERation Status Register
Bit 15
Bit 14
Bit 13
Bit 12
Bit 11 Bit 10
Bit 9
Bit 8
NU
NU
NU
NU
NU
NU
NU
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
NU
NU
NU
NU
NU
NU
NU
NU
∗
NU: not used