PSM SERIES PROGRAMMABLE POWER SUPPLY

PROGRAMMER MANUAL

⎯ 37 ⎯

:CURRent:PROTection:STATe

Function:
Set or query the overcurrent protection status.

Syntax:

:CURRent:PROTection:STATe {0|1|OFF|ON}

:CURRent:PROTection?

Examples:

:CURRent:PROTection:STATe 0

sets the over-current protection off.

If the overcurrent protection setting is on, the command of
CURRent:PROTection?

will return the value of 1.

:CURRent:PROTection:TRIPped?

Function:
Query if the overcurrent protection circuit is tripped.

Syntax:

:CURRent:PROTection:TRIPped?

Returns:

0|1

Examples:
This command will returns “1” if the overcurrent protection is tripped and
not cleared or “0” if not tripped.

:CURRent:PROTection:DELay

Function:
Set the OCP delay value.

Syntax:

:CURRent:PROTection:DELay{|MIN|MAX}

PSM SERIES PROGRAMMABLE POWER SUPPLY

PROGRAMMER MANUAL

⎯ 38 ⎯

Examples:
:CURRent:PROTection:DELay 3

sets the OCP delay value to 3 sec.

:CURRent:PROTection:DELay MIN

sets the OCP delay value to 0.1

sec.
:CURRent:PROTection:DELay MAX

sets the OCP delay value to 10.0

sec.

:CURRent:PROTection:DELay?

Function:
Query the OCP delay value.

Syntax:

:CURRent:PROTection:DELay?[MIN|MAX]

Return:

Examples:
:CURRent:PROTection:DELay?

will return the OCP value.

:CURRent:PROTection:DELay?

MIN will return 0.1 sec.

:CURRent:PROTection:DELay?

MAX will return 10.1 sec.

:VOLTage:PROTection

Function:
Set the overvoltage protection value.

Syntax:

:PROTection:VOLTage {|MIN|MAX}

Examples:

:VOLTage:PROTection

10 sets the overvoltage protection value to

10.000V(PSM-2010).

:VOLTage:PROTection MIN

sets the overvoltage protection value to

0.000V(PSM-2010).