PSM SERIES PROGRAMMABLE POWER SUPPLY

PROGRAMMER MANUAL

⎯ 35 ⎯

6)

:VOLTage:TRIGgered? MIN returns the lowest programmable

triggered voltage level.

:MEASure:CURRent?

Function:
Read the actual output current.

Syntax:

:MEASure:CURRent?

Returns:

Scientific mark

Examples:

:MEASure:CURRent?

might return +1.23400000E+00 to indicate that

the load is drawing 1.2340 A(PSM-2010).

:MEASure?

Function:
Read the actual output voltage.

Syntax:

:MEASure?

Returns:

Scientific mark

Examples:

:MEASure?

might return +1.15500000E+01 to indicate the voltage output

is 11.5500V(PSM-2010).

PSM SERIES PROGRAMMABLE POWER SUPPLY

PROGRAMMER MANUAL

⎯ 36 ⎯

:CURRent:PROTection

Function:
Set the overcurrent protection value.

Syntax:

:PROTection:CURRent {|MIN|MAX}

Examples:

:CURRent:PROTection

10 sets the overcurrent protection value to

10.000A(PSM-2010).

:CURRent:PROTection MIN

sets the overcurrent protection value to

0.000A(PSM-2010).

:CURRent:PROTection MAX

sets the overcurrent protection value to

22.000A(PSM-2010). Please refer to Table1-1, Table1-2 and Table1-3.

:CURRent:PROTection?

Function:
Query the overcurrent protection value.

Syntax:

:CURRent:PROTection? [MIN|MAX]

Returns:

Scientific mark

Examples:

:CURRent:PROTection?

will return the overcurrent protection value.

:CURRent:PROTection? MIN

will return 0.

:CURRent:PROTection? MAX

will return the highest programmable current

values. Please refer to Table1-1, Table1-2 and Table1-3.