GW Instek PSM-Series Programming Manual User Manual
Page 15

PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
⎯ 27 ⎯
*TST?
Function:
Self-test and test the RAM, ROM.
Syntax:
*TST?
Returns:
0|-300
Examples:
*TST?
returns 0, if the test is successful.
*TST?
returns –300, if the test is unsuccessful.
*WAI
Function:
WAI prevents the programming instrument from executing further
commands or queries until all pending operations are finished.
Syntax:
*WAI
z GENERAL SETTING COMMANDS
:APPLy
Function:
Set the output voltage and current value.
Syntax:
:APPLy {
:APPLy ?
PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
⎯ 28 ⎯
Examples:
:APPLy DEF, MAX
You can use DEF, MIN or MAX to replace voltage value <
Voltage
> or
current value
.
1) When select DEF, the voltage value is at 0, and the current value is at
20A(PSM-2010). Please refer to Table1-1, Table1-2 and Table1-3.
2) When select MIN, both values of voltage and current are at 0.
3) When select MAX, both values of voltage and current can be set at the
maximum value of the selected range. Please refer to Table1-1, Table1-
2 and Table1-3.
4) If only one parameter is specified, the power supply will regard it as
voltage value.
Ex
:APPLy 2.0
5)
:APPLy ?
This command returns present voltage and current setting
values.
0-8V/20A Range
0-20V/10A Range
Programming Range
0V to 8.24V
0V to 20.6V
MAX Value
8.24V
20.6V
MIN Value
0V
0V
DEFault Value
0V
0V
*RST Value
0V
MAX Ovp Value
22V
Voltage
MIN Ovp Value
0V
Programming Range
0A to 20.6A
0A to 10.3A
MAX Value
20.6A
10.3A
MIN Value
0A
0A
DEFault Value
20A
10A
*RST Value
20A
MAX Ocp Value
22A
Current
MIN Ocp Value
0A
Table1-1. PSM-2010 Programming Ranges