beautypg.com

GW Instek PSM-Series Programming Manual User Manual

Page 29

background image

PSM SERIES PROGRAMMABLE POWER SUPPLY

PROGRAMMER MANUAL

⎯ 55 ⎯

CURR:TRIG 3.0

TRIG:DEL 3.0

INIT

*TRG

After proceeding above program, the output will be change to 5V/3A
immediately.

z

CALIBRATION COMMANDS

CALibration:PASSword

Function:
Input Password to enter calibration mode.

Syntax:

CALibration:PASSword

The Password for PSM-2010 is 2010.

The Password for PSM-3004 is 3004.

The Password for PSM-6003 is 6003.

Examples:

CALibration:PASSword 2010

for PSM-2010 to enter calibration

mode

CALibration:TYPe

Function:
Select calibration mode.

Syntax:

CALibration:TYPe {0|1|2|3}

0:Voltage, 1:OVP , 2:Current 3:Ocp

PSM SERIES PROGRAMMABLE POWER SUPPLY

PROGRAMMER MANUAL

⎯ 56 ⎯

CALibration:VOLTage:MINimum

CALibration:VOLTage:MIDDle

CALibration:VOLTage:MAXimum

Function:
Input the voltage value separately.

Syntax:

CALibration:VOLTage:MINimum

CALibration:VOLTage:MIDDle

CALibration:VOLTage:MAXimum

CALibration:VOLTage:PROTection

Function:
Auto calibrate OVP.

Syntax:

CALibration:VOLTage:PROTection

CALibration:CURRent:MINimum

CALibration:CURRent:MIDDle

CALibration:CURRent:MAXimum

Function:
Input current value separately.