5. APPLICATION

5-1 The Table of the error message

Code

Cause

Resolvent

Err-01

Input voltage over 60V

Lower the input voltage

Err-02

Input power exceeds setting

Set higher power value or lower the voltage and

current

Err-03

Input voltage and power exceed setting

Lower the input voltage and set a higher power

value

Err-04

Input current exceeds setting

Lower the input current

Err-05

Input voltage and current exceed setting

Same as Err-01 and Err-04

Err-06

Input current and power exceed setting

Same as Err-02 and Err-04

Err-07

Input voltage, current and power exceed setting Same as ERR-01, Err-02 and Err-04

Err-08

Error of the adjusting data when performing

CHECKSUM

Send the instrument back to the manufacturer

for adjustment or adjust by yourself.

Err-09

Error inaccurate when performing CHECKSUM

Send the instrument back to the manufacturer

for repair

Err-10

The time of running AUTO has not been set

Re-set the time of execution

Err-11

Key-in value exceeds input range

Key in the value again

Err-12

The setting of STEP exceeds the resolution of

the range between chosen

Set the value of STEP again

Err-13

The setting of STEP exceeds output range

Set the value of V.I.R. again

To clear the Error Message: [SHIFT] + [C.C. MODE]