q When MEMORY indicates “01” and A indicates “CL03”, key in the current reading (no less than 6 digits

including decimal) of the DMM (in unit of “A”) and press

↵ .

r When MEMORY indicates “02”, key in the current reading (no less than 5 digits including decimal) of the

DMM (in unit of “A”) and press

↵ .

s When MEMORY indicates “03”, key in the current reading (no less than 5 digits including decimal) of the

DMM (in unit of “A”) and press

↵ .

t When MEMORY indicates “04”, key in the current reading (no less than 5 digits including decimal) of the

DMM (in unit of “A”) and press

↵ .

Note: Applied instruments -- POWER SUPPLY (18V, 20A); DMM (5 1/2 digits)

Applied situation --The C.R. MODE in 10

Ω range is not reliable.

4-4 Calibrating the OFFSET of C.R. MODE

n Set voltage of POWER SUPPLY to 3V and current of it to 0.5A
o Connect the output terminal of POWER SUPPLY to a DMM in 2A range in series. Then hook up the DMM

with the input terminal of the electronic load.

p Test the input terminal of the load by the 20V range of a DMM. Adjust the POWER SUPPLY until the

DMM display reads 3.00V.

q Set C.R. MODE to 10Ω (nselect C.R. MODE; o press V.I.R. SET; p key in “1”, “0”; q press ↵ ) first,

then set the load to LOAD ON status.

r adjust the POWER SUPPLY until the DMM display reads 0.300A.

Note: Applied instruments -- POWER SUPPLY (60V, 6A); DMM (5 1/2 digits)