beautypg.com

15 calibrating the d/a in 1 ω range of c.r. mode – GW Instek PEL-300 User Manual

Page 32

background image

29

4-15 Calibrating the D/A in 1

range of C.R. MODE

n Set voltage of POWER SUPPLY to 5V and current of it to 75A.
o Connect the output terminal of POWER SUPPLY to a DMM in 75A range in series. Then hook up the

DMM with the input terminal of the electronic load.

p Test the input terminal of the load by the 20V range of a DMM. The readout should be 5.00V.
q Press SHIFT and C.V. MODE, then key in 3, 8, 1,4, and ↵ . Then adjust the POWER SUPPLY until the

DMM display reads 5.00V.

r When MEMORY indicates “01” and A indicates “CL14”, key in the current reading (no less than 5 digits

including decimal) of the DMM (in unit of “A”) and press

↵ .

s When MEMORY indicates “02”, key in the current reading (no less than 5 digits including decimal) of the

DMM (in unit of “A”) and press

↵ .

t When MEMORY indicates “03”, key in the current reading (no less than 5 digits including decimal) of the

DMM (in unit of “A”) and press

↵ .

u When MEMORY indicates “04”, key in the current reading (no less than 5 digits including decimal) of the

DMM (in unit of “A”) and press

↵ .

Note: Applied instruments -- POWER SUPPLY (8V, 75A); DMM (5 1/2 digits)

Applied situation -- The readings of C.R. MODE are not reliable.