(1) Constant Voltage Mode:

n Connect the instrument according to Graph 3-3.
o Turn the power switch to “ON” position, then press C.V. MODE button (the light of C.V. MODE will be

on).

p Press LOAD ON/OFF button (the light of LOAD will be on), the displays will indicate the value of

external voltage and current.

Example:

To set input voltage at 20V in constant voltage mode

STEP 1.

Press C.V. MODE

STEP 2.

Press V.I.R. SET

STEP 3.

Press 20 (unit is specified as “V” )

STEP 4.

Press

(2). Constant Current Mode:

n Connect the instrument according to Graph 3-3.
o Turn the power switch to “ON” position, then press C.C. MODE button (the light of C.C. MODE is on).
p Press LOAD ON/OFF button (the light of LOAD is on), the displays will indicate the value of external

voltage and current.

Example:

To set input current at 1.248A in constant current mode

STEP 1.

Press C.C. MODE

STEP 2.

Press V.I.R. SET

STEP 3.

Press 1.248 (unit is specified as “A” )

STEP 4.

Press